TJ-CEO 2017 , Vol 12 , Num 4
Turkish Abstract Abstract PDF Similar Articles Mail to Author
Comparison of Anterior Segment Parameters Obtained with Two Different Scheimpfl ug Based Devices
Mete GÜLER1, Rumeysa TANYILDIZI2, Elif YUSUFOĞLU2, Nagehan CAN2, Fatih ÇELİK2
1Doç. Dr., Elazığ Eğitim ve Araştırma Hastanesi, Göz Hastalıkları Kliniği, Elazığ - Türkiye
2Uz. Dr., Elazığ Eğitim ve Araştırma Hastanesi, Göz Hastalıkları Kliniği, Elazığ - Türkiye
Purpose: To compare the corneal and anterior segment parameters obtained by Pentacam and Nidek AL-Scan in healthy subjects.

Material and Methods: The right eyes of 156 healthy volunteers were included in the study. Participants did not have a history of drug use and systemic illness and ocular surgery. Patients with 3.0 D or more astigmatism or 4.0 D or more spherical refractive error and history of previous contact lens use (within 4 weeks for rigid contact lenses and within 2 weeks for soft contact lenses) were also excluded. Anterior chamber depth (ACD), central corneal thickness (CCT), keratometry (K) [Flattest K (Kf), steepest K (Ks) and mean K (Km)] measurements obtained with Pentacam and Nidek AL-Scan optical biometer were compaired.

Results: Nidek AL Scan underestimated CCT measurements compared with Pentacam (p<0,001 and mean difference was -13.41 micrometers). Pearson correlation analysis determined signifi cant positive correlations between Nidek AL Scan and Pentacam in terms of CCT (r=0.981, p<0.001), ACD (r=0.983, p<0.001), Kf (r=0.990,p<0.001), Ks (r=0.990, p<0.001) and Km (r=0.991, p<0.001) The Bland-Altman plots showed narrow limits of agreement for ACD, Kf, Ks and Km parameters. However, the limits of agreement for CCT were broad and moderate agreement was found for this parameter.

Conclusions: Good agreement was found in Ks, Kf, Km and ACD parameters between Nidek AL Scan and Pentacam. On the other hand these devices do not seem to be interchangeable for pachymetric determination. Keywords : Anterior chamber depth, central corneal thickness; keratometry, nidek AL scan optical biometry, pentacam, scheimpflug