TJ-CEO 2006 , Vol 1 , Num 3
Turkish Abstract Abstract PDF Similar Articles Mail to Author
Comparison of Intraocular Pressure by Pascal Dynamic Contour Tonometry, Goldmann Applanation Tonometry, Non-contact Tonometry and Tono-pen and Effect of Central Corneal Thickness
Faruk ÖZTÜRK1, Tuncay KÜSBECİ2, Güliz YAVAŞ2, S. Samet ERMİŞ3, Ümit KAPLAN4, Ümit Übeyt İNAN3
1Afyon Kocatepe Üniversitesi Göz Hastalıkları A.D., Afyon, Prof. Dr.
2Afyon Kocatepe Üniversitesi Göz Hastalıkları A.D., Afyon, Yard. Doç. Dr.
3Afyon Kocatepe Üniversitesi Göz Hastalıkları A.D., Afyon, Doç. Dr.
4Afyon Kocatepe Üniversitesi Göz Hastalıkları A.D., Afyon, Dr.
Purpose: To compare the intraocular pressure (IOP) measurements obtained using Pascal Dynamic Contour Tonometer (PDCT), Goldmann Applanation tonometer (GAT), Non-contact tonometer (NCT) and Tono-pen and to correlate these with central corneal thickness (CCT) in eyes of normal healthy subjects.

Materials and Methods: IOP was measured with PDCT (SMT Swiss Microtechnology, Switzerland), GAT (Nikon, Japan), NCT ( Topcon CT- 80, Japan) and tono-pen (Tono-Pen XL, Medtronic Solan, USA) in random order in 106 eyes with normal corneas. CCT was measured using an ultrasonic pachymeter (Quentel Medical Clermont-Fernand, France). All measurements were repeated three times for each instrument. Mean values were evaluated with ANOVA and Pearson correlation analysis.

Results: Mean IOP was 16.1±2.7 mmHg, 14.0±2.7, 14.3±3.0 and 14.5±3.0 mmHg for PDCT, GAT, NCT and tono-pen, respectively. IOP values measured by PDCT were higher 2.0±2.5 mmHg, 1.7±2.7 mmHg, 1.5±2.4 mmHg than GAT, NCT and tono-pen values, respectively, and differences were found statistically significant (p<0.01, ANOVA). Mean CCT was 530.9±33 micrometer. PDCT, GAT and NCT values correlated with CCT (p<0.05, Pearson correlation analysis). Tono-pen showed no significant correlation with CCT (p>0.05, Pearson correlation analysis ).

Conclusions: We detected that IOP values measured by PDCT were higher than other methods. PDCT, GAT and NCT, except tono-pen, are affected by CCT in eyes with normal corneas of healthy subjects. Keywords : Pascal Dynamic Contour Tonometer , Goldmann Applanation Tonometer , Non-Contact Tonometer, Tono-pen, Central Corneal Thickness