ISSN : 3023-7505
  • Home
  • About The Journal
  • Editorial Board
  • Instructions for Authors
  • Contact
Current Issue
Ahead Of Print
Archive
Search
Most Popular
Download Articles Read Articles
Automated Perimetry
Carbonic Anhydrase Inhibitors
Intra Ocular Lens Power Calculation and Optic Biometry...
Visual Field Defects in Glaucoma
Visual Field Defect and Retinal Nerve Fiber Layer Defect in a Case of Optic Nerve Head Drusen...
Current Minimal Invasive Angle Procedures Without Implants for the Treatment of Glaucoma...
Intra Ocular Lens Power Calculation and Optic Biometry...
Automated Perimetry
Carbonic Anhydrase Inhibitors
Visual Field Defect and Retinal Nerve Fiber Layer Defect in a Case of Optic Nerve Head Drusen...
PureSee Kesintisiz Yüksek Kalitede Görüş
TJ-CEO 2014 , Vol 9 , Num 2
Turkish Abstract Abstract PDF Similar Articles Mail to Author
Analysis of Keratometry, Astigmatism and Axis Location Values with Different Instruments in Cataracts Patients
Mustafa ATAŞ1, Altan GÖKTAŞ2, Süleyman DEMİRCAN1, Emine PANGAL1, Ahmet GÜLHAN3, Gökmen ZARARSIZ4
1M.D. Kayseri Training and Research Hospital, Eye Clinic, Kayseri/TURKEY
2M.D. Associate Professor, Kayseri Training and Research Hospital, Eye Clinic, Kayseri/TURKEY
3M.D. Asistant, Kayseri Training and Research Hospital, Eye Clinic, Kayseri/TURKEY
4M.D. Asistant, Erciyes University Faculty of Medicine, Department of Biostatisitc, Kayseri/TURKEY
ABSTRACT Purpose: To compare keratometry, astigmatism and axis values with different instruments in cataract patients and to assess the agreement and correlation between these devices.

Materials and Methods: 62 eyes of 62 cataracts patients (37 men, 25 women) with a mean age of 64±11 years (41-84) were included in the study. Keratometry, astigmatism, axis values of the patients were analysed with autokeratometer, IOL master and Pentacam and The difference, agreement and correlation between the devices were compared statistically.

Results: Keratometry values were 43.41±1.58 D (39.75-47.25D), 43.54±1.56 D (39.8-47.44 D), and 43.25±1.61 D (39.10-47.10 D), astigmatism values were 0.93±0.93 D (0-4.50 D), 1,10 ±1.08 D (0.23-5.57 D), 0.96±1.00D (0.10-4.40 D) and axis values were 78.250±61.760 (0-1800), 96.290±59.360 (5-1800), 86.220±56.170 (1.00-1750) for autokeratometer, IOL master and Pentacam, respectively. Keratometry and astigmatism values measured with these 3 devices showed good agreement, correlation and repeatability (interclass correlation coefficients for all the values above 0.90). However, the relationship in axial location values were found to be weaker (interclass correlation coefficients values below 0.90).

Conclusion: Keratometry and astigmatism measured with these three devices show good agreement, correlation and repeatbility. However, axis location values were weaker compared to other measurements. Therefore, the axis values measured with these devices should not be used interchangably. Keywords :