TJ-CEO 2013 , Vol 8 , Num 4
Turkish Abstract Abstract PDF Similar Articles Mail to Author
The Comparison of Dynamic Contour Tonometry with Goldmann Applanation Tonometry, Non-Contact Tonometry and Tono-Pen; and the Association of Measurements with Ocular Structures
Ebru Nevin ÇETİN1, Kemal YAYLA2, Derya ÖZTUNA3, Avni AVUNDUK4, Volkan YAYLALI4, Cem YILDIRIM4
1M.D. Asistant Professor, Pamukkale University Faculty of Medicine, Department Ophthalmology, Denizli/TURKEY
2M.D., Pamukkale University Faculty of Medicine, Department Ophthalmology, Denizli/TURKEY
3M.D. Asistant Professor, Ankara University Faculty of Medicine, Department Bioistatistic, Ankara/TURKEY
4M.D. Professor, Pamukkale University Faculty of Medicine, Department Ophthalmology, Denizli/TURKEY
Purpose: To compare the intraocular pressure (IOP) measured by dynamic contour tonometry (DCT), Goldmann applanation tonometry (GAT), non-contact tonometry (NCT) and Tono-Pen and also to assess whether ocular parameters such as corneal thickness, keratometry and axial length affect the measurements by the tonometers.

Materials and Methods: One hundred healthy subjects prospectively underwent IOP measurement with DCT, GAT, NCT and Tono-Pen. The correlation between the tonometer readings as well as the effect of keratometry, central corneal thickness (CCT), axial length and anterior chamber depth on IOP measurements were assessed.

Results: DCT readings were higher than NCT (ΔIOP=2.0±3.2 mmHg, p<0.001) and GAT (ΔIOP=1.3±2.1 mmHg, p=0.003) but not Tono-Pen (ΔIOP=0.1±2.7 mmHg, p=0.981). DCT, GAT and NCT correlated positively with CCT (p=0.039, p=0.020 and p=0.001). GAT and NCT readings were significantly higher in eyes with thick cornea (>568 μm) compared to eyes with thin cornea (<535 μm) (p=0.021 and p=0.003). Ocular parameters other than CCT did not have a significant effect on the readings of tonometers.

Conclusion: Despite significant correlation; DCT readings were higher than GAT and NCT but not Tono-Pen. DCT, NCT and GAT readings significantly correlated with CCT. GAT and NCT readings were significantly higher in eyes with thick cornea compared to eyes with thin cornea. Keywords : Dynamic contour tonometry, Goldmann applanation tonometry, non-contact tonometry, pascal tonometry, Tono-Pen