TJ-CEO 2012 , Vol 7 , Num 2
Turkish Abstract Abstract PDF Similar Articles Mail to Author
Comparison of Goldmann Applanation Tonometry, Tonopen-Avia and Ocular Response Analyser On Intraocular Pressure Measurement and The Effect of Central Corneal Thickness
Işıl PAŞAOĞLU1, Mümin Hakan EREN1, Ali DEMIRCAN2, Hülya GÜNGEL3, Çiğdem ALTAN1, Ercan PAŞAOĞLU4
1M.D., Beyoglu Training and Research Hospital, Eye Clinic, İstanbul/TURKEY
2M.D., Beyoglu Training and Research Hospital, Eye Clinic, İstanbul/TURKEY
3M.D. Professor, Beyoglu Training and Research Hospital, Eye Clinic, İstanbul/TURKEY
4M.D., Necmi Ayanoglu Silivri State Hospital, Eye Clinic, İstanbul/TURKEY
Purpose: To compare the intraocular pressure measurements obtained with a Goldmann applanation tonometer (GAT), Tono-Pen AVIA, and ocular response analyzer (ORA) and to investigate the effect of central corneal thickness (CCT).
Materials and Methods: Seventy-seven eyes followed because of primary open angle glaucoma were studied prospectively. IOP was measured by GAT, Tono-Pen AVIA, and ORA. The patients were divided into 3 groups as follows: group 1, thin (<545 µm); group 2, moderate (545-580 µm); and group 3, thick (>580 µm) according to CCT measured by ultrasonic pachymetry. Measurements were compared between the groups.
Results: The average age was 57.31±10.87 years. The average CCT was 528.92±12.58 μm for group 1, 557.40±9.14 μm for group 2, and 601.90±19.01 μm for group 3. GAT, Tono-Pen AVIA, and ORA measurements were 15.6±3.3, 16.7±2.9, and 16.9±3.9 mmHg in the whole study group; 15.1±3.9, 16.2±3.0, and 16.9±4.4 mmHg in group 1; 15.4±2.8, 16.5±2.5, and 16.4±3.7 mmHg in group 2; and 16.4±3.1, 17.7±3.4, 17.6± 3.6 mmHg in group 3, respectively. GAT measurements were significantly lower than Tono-Pen AVIA and ORA measurements in the whole study group (p=0.001, p=0.000), but there was no significant difference between Tono-Pen AVIA and ORA (p=0.64). According to CCT, while GAT values were lower than Tono-Pen AVIA measurements in groups 2 and 3 (p=0.015, p=0.18), no significant difference was observed in group 1 (p=0.075). Moreover, GAT values were lower than ORA measurements in groups 1 and 2 (p=0.003 and p=0.020), but there was no significant difference in group 3 (p=0.068).
Conclusion: IOP values measured with GAT were lower than those measured with ORA in thin corneas, and lower than those measured with Tono-Pen AVIA in thick corneas.
Keywords : Intraocular pressure, Goldmann applanation tonometer, Tono-Pen AVIA, ocular response analyzer, central corneal thickness